CFP last date
22 April 2024
Reseach Article

Practical Assessment of MRI Coil's Characteristics

by Dia M. Ali
International Journal of Computer Applications
Foundation of Computer Science (FCS), NY, USA
Volume 179 - Number 17
Year of Publication: 2018
Authors: Dia M. Ali
10.5120/ijca2018916287

Dia M. Ali . Practical Assessment of MRI Coil's Characteristics. International Journal of Computer Applications. 179, 17 ( Feb 2018), 27-30. DOI=10.5120/ijca2018916287

@article{ 10.5120/ijca2018916287,
author = { Dia M. Ali },
title = { Practical Assessment of MRI Coil's Characteristics },
journal = { International Journal of Computer Applications },
issue_date = { Feb 2018 },
volume = { 179 },
number = { 17 },
month = { Feb },
year = { 2018 },
issn = { 0975-8887 },
pages = { 27-30 },
numpages = {9},
url = { https://ijcaonline.org/archives/volume179/number17/28960-2018916287/ },
doi = { 10.5120/ijca2018916287 },
publisher = {Foundation of Computer Science (FCS), NY, USA},
address = {New York, USA}
}
%0 Journal Article
%1 2024-02-07T00:55:39.225420+05:30
%A Dia M. Ali
%T Practical Assessment of MRI Coil's Characteristics
%J International Journal of Computer Applications
%@ 0975-8887
%V 179
%N 17
%P 27-30
%D 2018
%I Foundation of Computer Science (FCS), NY, USA
Abstract

Magnetic Resonance coil is one of the limitation and degradation of image quality in Medical Imaging. Ghosting, Artifacts, and noise are the main consequence of the mismatch of the receiving coil. Human body parts have deferent permittivity and permeability which cause a variation of the self-resonance inductance. I this paper, the effects of variable material filled in brain coil is studied and measured. The different arrangement also performed. The results show a noticeable variation of the self-resonance inductance which yields drifting the operating frequency by 122 KHz which is beyond the bandwidth of the MRI signal and instrument. One of the solutions is the auto-tuning algorithm, which is the only available solution. As a proposed solution, a cylindrical slab with high relative permittivity can be used inside the coil winding which reduce the variation of the operating frequency..

References
  1. J. Vaughan, M. Garwood, et. al “7T vs. 4T: RF power, homogeneity, and signal-to-noise comparison in head images,” Magnetic Resonance in Medicine, vol. 46, no. 1, pp. 24–30, 2001.
  2. ]J. T. Vaughan, C. J. Snyder, L. J. DelaBarre, P. J. Bolan, J. Tian, L. Bolinger, G. Adriany, P. Andersen, J. Strupp, and K. Ugurbil, “Whole-body imaging at 7T: Preliminary results,” Magnetic Resonance in Medicine, vol. 61, no. 1, pp. 244–248, 2009..
  3. A. J. E. Raaijmakers, O. Ipek, D. W. J. Klomp, C. Possanzini, P. R. Harvey, J. J. W. Lagendijk, and C. a. T. van den Berg, “Design of a radiative surface coil array element at 7T: The single-side adapted dipole antenna,” Magnetic Resonance in Medicine, vol. 66, no. 5, pp. 1488–1497, 2011.
  4. C. E. Hayes, W. A. Edelstein, J. F. Schenck, O. M. Mueller &M. Eash (1985). An efficient, highly homogeneous radio frequency coil for whole-body NMR imaging at 1.5 T. J. Magn. Res., 63, 622–8.
  5. E.M. Haacke, R.W. Brown, M.R. Thompson, R. Venkatesan, Magnetic Resonance Imaging: Physical Principles and Sequence Design Book, Wiley-Liss, 1999.
  6. C.E. Hayes, W.A. Edelstein, J.F. Schenck, O.M. Mueller, M. Eash, “An efficient, highly homogeneous radio frequency coil for whole-body NMR imaging at 1.5 T”, J. Magn. Reson., vol. 63(3), pp. 622–628, 1985.
  7. P.B. Roemer, W.A. Edelstein, C.E. Hayes, S.P. Souza, O.M. Mueller, “The NMR phased array”, Magn. Reson. Med., vol.16, pp. 192–225, 1990.
  8. S.M. Wright, L.L. Wald, “Theory and application of array coils in MR spectroscopy”, NMR Biomed., vol. 10, pp. 394–410, 1997.
  9. Y. Wang, Z. Shi, Y. Ma, C. Wang, M. Chena nd Y. Chen, ‘Supercapacitor Devices Based on Graphene Materials’, J. Phys. Chem. C 113 , 13103 (2009).
  10. M. Jayalakshmi and K. Balasubramanian, “Review: Simple Capacitors to Supercapacitors,” International Journal of Electrochemical Science (2008)
Index Terms

Computer Science
Information Sciences

Keywords

MRI Receiving Coils Q Point..