Different Sizes of Windows for Energy Spectrum of a Photon Counting X-ray Imaging System

by Muhammad Masood Sarfaraz, Aamir Khan, Farah Batool, Adeel Akram, Farman Ullah
International Journal of Computer Applications
Foundation of Computer Science (FCS), NY, USA
Volume 39 - Number 10
Year of Publication: 2012
Authors: Muhammad Masood Sarfaraz, Aamir Khan, Farah Batool, Adeel Akram, Farman Ullah
10.5120/4854-6946

Muhammad Masood Sarfaraz, Aamir Khan, Farah Batool, Adeel Akram, Farman Ullah . Different Sizes of Windows for Energy Spectrum of a Photon Counting X-ray Imaging System. International Journal of Computer Applications. 39, 10 ( February 2012), 8-13. DOI=10.5120/4854-6946

@article{ 10.5120/4854-6946,
author = { Muhammad Masood Sarfaraz, Aamir Khan, Farah Batool, Adeel Akram, Farman Ullah },
title = { Different Sizes of Windows for Energy Spectrum of a Photon Counting X-ray Imaging System },
journal = { International Journal of Computer Applications },
issue_date = { February 2012 },
volume = { 39 },
number = { 10 },
month = { February },
year = { 2012 },
issn = { 0975-8887 },
pages = { 8-13 },
numpages = {9},
url = { https://ijcaonline.org/archives/volume39/number10/4854-6946/ },
doi = { 10.5120/4854-6946 },
publisher = {Foundation of Computer Science (FCS), NY, USA},
address = {New York, USA}
}
%0 Journal Article
%1 2024-02-06T20:26:04.168314+05:30
%A Muhammad Masood Sarfaraz
%A Aamir Khan
%A Farah Batool
%A Adeel Akram
%A Farman Ullah
%T Different Sizes of Windows for Energy Spectrum of a Photon Counting X-ray Imaging System
%J International Journal of Computer Applications
%@ 0975-8887
%V 39
%N 10
%P 8-13
%D 2012
%I Foundation of Computer Science (FCS), NY, USA
Abstract

In relation to the radiation imaging sensor Timepix, the energy spectrum generated from the sensor data has been studied and for this, different sizes of windows have been selected in order to observe their effect on energy spectrum generation. It was found that removing the charge sharing effect significantly improves the means of discovering the centre peak. Additionally, it has been observed that a 4x4 window is best suited with regards to energy spectrum generation. Different fluorescent materials were used as samples in order to generate test images and the analysis was conducted using MATLAB.

References
  1. Victor Gutierrez Diez,’Measurements of time pix detectors performance’, Universidad Complutense de Madrid, September 2010.
  2. http://aladdin.utef.cvut.cz/ofat/Methods/TimePixCalibration/index.html
  3. Erik Fröjdh,’X-ray absorption and charge transport in a pixellatedCdTe detector with single photon processing read out’, Mid Sweden University , Holmgatan 10,851 70 Sundsvall, Sweden .
  4. X. Llopart,’ Timepix a 65K programmable pixel readout chip for arrival time, energy and/or photon counting measurements’, Volume 581, Issues 1-2,21 October 2007 , pages 485-494. VCL 2007-Proceedings of the 11th International Vienna Conference on Instrumentation.
  5. Medipix in IEAP, available at: http://aladdin.utef.cvut.cz/ofat/Methods/TimePixCalibration/index.htm#Clusters
  6. Fluorescence Tutorials, available at: http://www.invitrogen.com/site/us/en/home/support/Tutorials.html
  7. Glenn F. Knoll, ‘Radiation Detection and Measurement’, ISBN 0-471-07338-5
  8. Syed Naeem Ahmed, ‘Physics & Engineering of Radiation Detection by’, ISBN -10:0-12-045581-1
  9. Muhammad MasoodSarfaraz, “An Investigation of How Pixel Size Affects the Energy Resolution of a Photon Counting X-ray Imaging System”, MS Thesis 2011
Index Terms

Computer Science
Information Sciences

Keywords

Radiation Imaging sensor Timepix Energy Spectrum